Muffle furnace is commonly used in hot processing and industrial workpiece processing equipment, because of reasonable design, simple operation, to the heat treatment work has brought great convenience, and the important thing in Muffle furnace is whether the temperature is fine. Indeed, this determines the quality of the workpiece produced by the Muffle furnace. The following Xiaobian will explain to you what factors affect the temperature error of the Muffle furnace.
Factors affecting temperature error of Muffle furnace

Error caused by thermocouples
The Seebeck effect is used to measure the temperature of a thermocouple. a,? b is connected into a closed loop when the two contact points are at t0. In this way, an electromotive force called the thermoelectric potential is generated in the loop, a phenomenon called thermoelectric action. The thermoelectric potential varies with temperature t0 and t, where, at constant contact temperature t0, the thermoelectric potential is temperature. A single value function of t, by measuring the magnitude of the thermal potential, can find the measured temperature t. Causes of thermocouple error are:
1, the thermocouple is not uniform
If the thermocouple is made of homogeneous conductors, its thermoelectric potential is only related to the temperature at its ends. If the thermal electrode material is not uniform, and the thermal electrode is in a temperature gradient field, then the thermocouple will generate an additional heating potential, the size of which is determined by the temperature difference ladder distribution along the length of the thermal electrode. The non-uniformity of the material is related to the degree of non-uniformity and the position of the hot spot potential in the temperature field.
2, the thermocouple is unstable
Instability refers to the change of the thermocouple score with the use time and conditions of the index value. Many times, this can be a major factor in temperature errors.
Muffle furnace
Display error
All display instruments have an accuracy class, which indicates the tolerance for error in the second measurement. The reasons for this error are:
1. The influence of cold end temperature
The thermoelectric potential generated by the thermocouple, in addition to the temperature of the measurement end, is also related to the temperature of the cold end, and only when the cold end temperature is constant, the thermoelectric potential is only related to the temperature of the working end. Most of the instruments have the cold end temperature compensation function, when the instrument is running, the change of the ambient temperature is small, the compensation effect on the cold end temperature is minimal, but when the ambient temperature changes greatly, because the instrument can not fully compensate the cold end temperature, it will also produce some errors. Where are you? In type b thermocouples, when the thermoelectric potential of 0~50 ° C is less than 3 microvolts, temperature compensation is not required.
2. Nonlinear influence
It can be seen from the characteristic curve of the forming thermocouple that there is no direct relationship between the thermoelectric potential and temperature of the two different materials composed of the thermocouple, and even some temperature differences are large. Therefore, when using thermocouples for temperature measurement, nonlinear correction must be carried out, otherwise it will cause errors.
3, the impact of signal processing
From the index table of the thermocouple, it can be seen that the thermoelectric motive force of the thermocouple is very small, and to achieve higher accuracy, the precision of the measuring instrument must be increased and the external circuit must be enlarged, which will also bring errors.